DC/DC Medical Converter
•
•
•
I/O isolation 5000 VAC (reinforced)
•
Certification according to IEC/EN/ES
60601-1 3rd edition for 2xMOPP and
IEC/EN/UL 62368-1
•
•
•
•
•
•
TRV 2M Series, 2 Watt
Wide 1.5:1 input voltage range
Operating temperature range
-40 to +80 °C without derating
Short circuit protection
Regulated outputs
Low leakage current < 2 μA
Efficiency up to 84 %
Operation up to 5000 m altitude
5-year product warranty
ES 60601-1 IEC 60601-1
UL 62368-1 IEC 62368-1
The TRV 2M is a series of 2 Watt DC/DC converters in a compact SIP-9 package with reinforced isolation of 5000 VAC for medical and industrial applications. The series offers a 1.5:1 input voltage range with a nominal input ranging
between 5 and 24 VDC. With a continuous short circuit protection and a low
leakage current of less than 2 µA, this converter is especially suited to protect
any connected interfaces or applied parts to patients. Featuring almost fully
regulated outputs this series provides great level of regulation without affecting
the cost efficiency. It is an ideal solution for applications where an unregulated
DC/DC converter would not meet your regulation requirements but cost still is a
critical factor. With an operating temperature range from -40 to +80°C without
derating and certifications acc. IEC/EN/ES 60601-1 3rd ed. for 2xMOPP and
IEC/EN/UL 62368-1 this series is suitable for many different applications
where a medical isolation system and short circuit protection is needed.
Models
Order Code
Input Voltage
Range
TRV 2-0510M
TRV 2-0511M
TRV 2-0512M
TRV 2-0513M
TRV 2-0521M
TRV 2-0522M
TRV 2-0523M
TRV 2-1210M
TRV 2-1211M
TRV 2-1212M
TRV 2-1213M
TRV 2-1221M
TRV 2-1222M
TRV 2-1223M
TRV 2-1510M
TRV 2-1511M
TRV 2-1512M
TRV 2-1513M
TRV 2-1521M
TRV 2-1522M
TRV 2-1523M
TRV 2-2410M
TRV 2-2411M
TRV 2-2412M
TRV 2-2413M
TRV 2-2421M
TRV 2-2422M
TRV 2-2423M
Note
4.5 - 7 VDC
(5 VDC nom.)
9.6 - 14.4 VDC
(12 VDC nom.)
12 - 18 VDC
(15 VDC nom.)
19.2 - 28.8 VDC
(24 VDC nom.)
Output 1
Output 2
Vnom
Imax
3.3 VDC
5 VDC
12 VDC
15 VDC
+5 VDC
+12 VDC
+15 VDC
3.3 VDC
5 VDC
12 VDC
15 VDC
+5 VDC
+12 VDC
+15 VDC
3.3 VDC
5 VDC
12 VDC
15 VDC
+5 VDC
+12 VDC
+15 VDC
3.3 VDC
5 VDC
12 VDC
15 VDC
+5 VDC
+12 VDC
+15 VDC
600 mA
400 mA
167 mA
134 mA
200 mA
83 mA
67 mA
600 mA
400 mA
167 mA
134 mA
200 mA
83 mA
67 mA
600 mA
400 mA
167 mA
134 mA
200 mA
83 mA
67 mA
600 mA
400 mA
167 mA
134 mA
200 mA
83 mA
67 mA
Vnom
Efficiency
Imax
–5 VDC
–12 VDC
–15 VDC
200 mA
83 mA
67 mA
–5 VDC
–12 VDC
–15 VDC
200 mA
83 mA
67 mA
–5 VDC
–12 VDC
–15 VDC
200 mA
83 mA
67 mA
–5 VDC
–12 VDC
–15 VDC
200 mA
83 mA
67 mA
typ.
78 %
81 %
83 %
83 %
82 %
83 %
81 %
79 %
81 %
84 %
83 %
81 %
83 %
82 %
79 %
81 %
84 %
83 %
81 %
83 %
80 %
78 %
80 %
82 %
82 %
81 %
81 %
80 %
- If the input will be switched electromechanically, use an external 47 μF/63 V E/C. to reduce voltage transient.
www.tracopower.com
September 20, 2023
Page 1 / 5
TRV 2M Series, 2 Watt
Input Specifications
Input Current
5 Vin models: 5 mA typ. (3.3 Vout model)
5 mA typ. (5 Vout model)
20 mA typ. (12 Vout model)
20 mA typ. (15 Vout model)
5 mA typ. (5 / -5 Vout model)
20 mA typ. (12 / -12 Vout model)
25 mA typ. (15 / -15 Vout model)
12 Vin models: 4 mA typ. (3.3 Vout model)
4 mA typ. (5 Vout model)
10 mA typ. (12 Vout model)
10 mA typ. (15 Vout model)
4 mA typ. (5 / -5 Vout model)
10 mA typ. (12 / -12 Vout model)
10 mA typ. (15 / -15 Vout model)
15 Vin models: 4 mA typ. (3.3 Vout model)
4 mA typ. (5 Vout model)
8 mA typ. (12 Vout model)
8 mA typ. (15 Vout model)
4 mA typ. (5 / -5 Vout model)
8 mA typ. (12 / -12 Vout model)
8 mA typ. (15 / -15 Vout model)
24 Vin models: 3 mA typ. (3.3 Vout model)
3 mA typ. (5 Vout model)
6 mA typ. (12 Vout model)
6 mA typ. (15 Vout model)
3 mA typ. (5 / -5 Vout model)
6 mA typ. (12 / -12 Vout model)
6 mA typ. (15 / -15 Vout model)
5 Vin models: 510 mA typ.
12 Vin models: 210 mA typ.
15 Vin models: 170 mA typ.
24 Vin models: 105 mA typ.
5 Vin models: 15 VDC max. (1 s max.)
12 Vin models: 25 VDC max. (1 s max.)
15 Vin models: 25 VDC max. (1 s max.)
24 Vin models: 35 VDC max. (1 s max.)
5 Vin models: 1'000 mA (slow blow)
12 Vin models: 500 mA (slow blow)
15 Vin models: 500 mA (slow blow)
24 Vin models: 315 mA (slow blow)
(The need of an external fuse has to be assessed
in the final application.)
Internal Capacitor
- At no load
- At full load
Surge Voltage
Recommended Input Fuse
Input Filter
Output Specifications
Voltage Set Accuracy
Regulation
- Input Variation (Vmin - Vmax)
- Load Variation
- Voltage Balance
(symmetrical load)
- Cross Regulation
(25% / 100% asym. load)
single output models:
dual output models:
See application note:
dual output models:
±3% max. (60% load: 3.3, 5, ±5 Vout models)
±3% max. (90% load: other models)
2% max.
2% max.
www.tracopower.com/overview/trv2m
4% max.
dual output models: 4% max.
All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.
www.tracopower.com
September 20, 2023
Page 2 / 5
TRV 2M Series, 2 Watt
Ripple and Noise
(20 MHz Bandwidth)
- single output
3.3 Vout models:
5 Vout models:
12 Vout models:
15 Vout models:
5 / -5 Vout models:
12 / -12 Vout models:
15 / -15 Vout models:
3.3 Vout models:
5 Vout models:
12 Vout models:
15 Vout models:
5 / -5 Vout models:
12 / -12 Vout models:
15 / -15 Vout models:
- dual output
Capacitive Load
- single output
- dual output
Minimum Load
Temperature Coefficient
Start-up Time
Short Circuit Protection
100 mVp-p typ.
100 mVp-p typ.
125 mVp-p typ.
125 mVp-p typ.
100 / 100 mVp-p typ.
125 / 125 mVp-p typ.
125 / 125 mVp-p typ.
2'000 µF max.
820 µF max.
470 µF max.
470 µF max.
470 / 470 µF max.
220 / 220 µF max.
220 / 220 µF max.
Not required
±0.03 %/K max.
20 ms max.
Continuous, Automatic recovery
Safety Specifications
Safety Standards
EN 62368-1
IEC 62368-1
UL 62368-1
EN 60601-1
IEC 60601-1
ANSI/AAMI ES 60601-1
2 x MOPP (Means Of Patient Protection)
www.tracopower.com/overview/trv2m
PD 2
Not mains connected
- IT / Multimedia Equipment
- Medical Equipment
- Certification Documents
Pollution Degree
Over Voltage Category
EMC Specifications
EMI Emissions
- Conducted Emissions
- Radiated Emissions
External filter proposal:
EMS Immunity
- Electrostatic Discharge
Air:
Contact:
- RF Electromagnetic Field
- EFT (Burst) / Surge
External filter proposal:
- Conducted RF Disturbances
- PF Magnetic Field
Continuous:
1 s:
EN 60601-1-2 edition 4 (Medical Devices)
EN 55011 class A (with external filter)
EN 55011 class B (with external filter)
EN 55032 class A (with external filter)
EN 55032 class B (with external filter)
EN 55011 class A (with external filter)
EN 55011 class B (with external filter)
EN 55032 class A (with external filter)
EN 55032 class B (with external filter)
www.tracopower.com/overview/trv2m
EN 55024 (IT Equipment)
EN 55035 (Multimedia)
EN 60601-1-2 edition 4 (Medical Devices)
EN 61000-4-2, ±15 kV, perf. criteria A
EN 61000-4-2, ±8 kV, perf. criteria A
EN 61000-4-3, 10 V/m, perf. criteria A
EN 61000-4-4, ±2 kV, perf. criteria A
EN 61000-4-5, ±2 kV, perf. criteria A
www.tracopower.com/overview/trv2m
EN 61000-4-6, 10 Vrms, perf. criteria A
EN 61000-4-8, 100 A/m, perf. criteria A
EN 61000-4-8, 1000 A/m, perf. criteria A
General Specifications
Relative Humidity
95% max. (non condensing)
All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.
www.tracopower.com
September 20, 2023
Page 3 / 5
TRV 2M Series, 2 Watt
Temperature Ranges
Power Derating
Cooling System
Altitude During Operation
Switching Frequency
Insulation System
Working Voltage (rated)
Isolation Test Voltage
Creepage
Clearance
Isolation Resistance
Isolation Capacitance
- Operating Temperature
- Case Temperature
- Storage Temperature
- High Temperature
- Input to Output, 60 s
- Input to Output, 1 s
- Input to Output
- Input to Output
- Input to Output, 500 VDC
- Input to Output, 100 kHz, 1 V
Leakage Current
Reliability
Washing Process
- Touch Current
- Calculated MTBF
Environment
- Vibration
- Mechanical Shock
- Thermal Shock
Housing Material
Potting Material
Pin Material
Pin Foundation Plating
Pin Surface Plating
Housing Type
Mounting Type
Connection Type
Footprint Type
Soldering Profile
Weight
Environmental Compliance - REACH Declaration
- RoHS Declaration
- SCIP Reference Number
–40°C to +95°C
+105°C max.
–55°C to +125°C
Depending on model
See application note: www.tracopower.com/overview/trv2m
Natural convection (20 LFM)
5'000 m max.
180 - 360 kHz (PFM)
Reinforced Insulation
250 VAC
5'000 VAC
7'000 VDC
8 mm min.
8 mm min.
10'000 MΩ min.
10 pF typ.
20 pF max.
2 µA max. (240 VAC, 60 Hz)
10'410'000 h (MIL-HDBK-217F, ground benign)
According to Cleaning Guideline
www.tracopower.com/info/cleaning.pdf
MIL-STD-810F
7.6 g, 3 axis, 60 min, 20-2000 Hz
MIL-STD-810F
50 g, 3 axis, 11 ms
MIL-STD-810F
-55°C to +125°C, 72 cycles, 30 min each
Non-conductive Plastic (UL 94 V-0 rated)
Silicone (UL 94 V-0 rated)
Tinned Copper
Nickel (2 - 3 µm)
Tin (3 - 5 µm), matte
Plastic Case
PCB Mount
THD (Through-Hole Device)
SIP9
Lead-Free Wave Soldering
260°C / 6 s max.
4.8 g
www.tracopower.com/info/reach-declaration.pdf
REACH SVHC list compliant
REACH Annex XVII compliant
www.tracopower.com/info/rohs-declaration.pdf
Exemptions: 7a, 7c-I
(RoHS exemptions refer to the component
concentration only, not to the overall
concentration in the product (O5A rule).)
2dda669e-cfb2-4463-b721-ad8acc4e0fbd
Supporting Documents
Overview Link (for additional Documents)
www.tracopower.com/overview/trv2m
All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.
www.tracopower.com
September 20, 2023
Page 4 / 5
TRV 2M Series, 2 Watt
Outline Dimensions
19.6
Pinout
2.0
1
+Vin
+Vin
2
in
in
7
out
out
8
No pin
Common
9
+Vout
+Vout
2.0
(0.07)
9.9
1.8
(0.63)
Bottom view
(0.02x0.01)
Dual
(0.079) (0.079)
16.0
0.5x0.25
Single
(0.39)
10.0
(0.394)
Pin
9
4.1
2.0
(0.079)
8
(0.16)
7
2.3
2
(0.09)
1
(0.49)
12.5
(0.77)
Dimensions in mm (inch)
Tolerances: x.x ±0.5 (±0.02)
x.xx ±0.25 (±0.01)
Pin diameter ±0.1 (±0.004)
© Copyright 2023 Traco Electronic AG
Specifications can be changed without notice.
www.tracopower.com
Rev. September 20, 2023
Page 5 / 5